Logo image
MAGNETICALLY INDUCED FORCE AND TORQUE ON TEMPORARY EPICARDIAL LEADS: EXPERIMENTAL DEVICE DEVELOPMENT AND CLINICAL TESTING
Thesis   Open access

MAGNETICALLY INDUCED FORCE AND TORQUE ON TEMPORARY EPICARDIAL LEADS: EXPERIMENTAL DEVICE DEVELOPMENT AND CLINICAL TESTING

Daniel Juarez Luna
Master of Science (MS), Creighton University
2026

Abstract

Epicardial pacing leads Magnetic torque quantification MRI safety Static magnetic field interactions Translational force measurement Medical Imaging
Patients with retained epicardial pacing leads following cardiac surgery are restricted MRI access due to risks of static-field translational forces and torques, despite frequent diagnostic need for MRI in this patient population. This thesis developed and clinically tested three custom non-magnetic fixtures to quantify these interactions at micro-Newton scales: a PVC platform for magnetic field characterization, a pendulum for displacement force, and a strain-gauge sensor and fixture for torque. Each fixture was validated in a controlled laboratory environment using an air-core solenoid and Helmholtz/neodymium fields (≤180 G), then tested on a 3 T Siemens Magnetom™ scanner at CHI Health CUMC–Bergan Mercy. Laboratory field mapping confirmed the expected field spatial decay, establishing the |B∇B| product needed for force predictions. Trial pendulum measurements of a ferromagnetic object yielded χm/ρ = (2.01 ± 0.05) × 10 3 m3/kg (tan α = 0.002–0.012), validating the linear model before clinical deployment. Torque sensor calibration produced consistent linear voltage-to-force slopes across all orientations (τmin = 1.24 × 10 5N· m), and laboratory measurements confirmed τ ∝ B2 dependence (k = 379 ± 9 N/m, 2.3%). Clinical field mapping yielded |B∇B| between 12.68 and 22.95 T2/m at the bore entrance. Translational-force ratios (FR = tan α) remained below the ASTM F2052-15 threshold of 1.0 for all tested materials. The Medtronic STREAMLINE 6492 unipolar epicardial lead reached the highest deflection (α = 9.5 , FR = 0.166), substantially higher than titanium references (FR = 0.006) yet well within the safety criterion. Extrapolation yielded allowable gradients of 107 T/m at 1.5 T and 54 T/mat 3.0 T for the unipolar lead, both far exceeding the scanner’s physical maximum (≤ 19 T/m). Clinical torque runs at B ≈ 1.5 T produced no measurable deflection, requiring further investigation. The results indicate that micro-Newton-scale magnetic-field mechanical effects on the tested epicardial lead can be measured under 3 T conditions. Though the findings are limited to a single configuration and do not support general safety conclusions, it established a methodology that can be extended to evaluate a broader range of leaddesigns and conditions.
pdf
JuarezLuna_D_2026_MS9.91 MBDownloadView
Open Access

Metrics

1 Record Views

Details

Logo image