Agreement Between the TEG 6s and TEG 5000 Analyzers in Extracorporeal Membrane Oxygenation
Journal article   Peer reviewed

Agreement Between the TEG 6s and TEG 5000 Analyzers in Extracorporeal Membrane Oxygenation

Abhimanyu Chandel, Mehul Desai, Christopher S. King, Saloni Patolia, Anika I. Raja, Ramesh Singh and Heidi J. Dalton
ASAIO Journal, Vol.68(1)
2022

Abstract

Adult Blood Coagulation Tests Extracorporeal Membrane Oxygenation Humans Reproducibility of Results Thrombelastography Thrombosis Respiratory therapy Vibration analysis Viscoelasticity heparin kaolin Blood clot formation Confidence interval Extracorporeal membrane oxygenation Haemostasis Operator error Performance Thromboelastography Vibration errors Viscoelastic properties Viscoelastic testing adult anticoagulant therapy Article extracorporeal oxygenation female hospital patient human information processing intermethod comparison male middle aged observational study reaction time adverse event blood clotting test reproducibility thromboelastography thrombosis Oxygenation
Thromboelastography (TEG) evaluates viscoelastic properties of blood clot formation. The TEG 5000 analyzer is commonly used but prone to errors related to vibration or operator error. The TEG 6s was developed to overcome these limitations. Performance of TEG 6s compared with TEG 5000 has not been reported in extracorporeal membrane oxygenation (ECMO). We compared the agreement between devices via a single-center prospective observational study in hospitalized adult patients on ECMO. Data for both devices were collected daily for 3 days after ECMO initiation. Standard tests for method comparison were used. Thirty-four matching samples were available for analysis. Minimal bias was noted; however, the limit of agreement was wide for TEG parameters. Visually, agreement was better for values within the reference ranges of the tests. Lin's coefficients demonstrated moderate correlation for reaction time and alpha angle (0.58; 95% confidence interval [CI], 0.31-0.76 and 0.63; 95% CI, 0.40-0.78, respectively). Excellent correlation was demonstrated for kinetic time and maximum amplitude (0.88; 95% CI, 0.79-0.94 and 0.89; 95% CI, 0.79-0.94). The TEG 6s device may represent an acceptable surrogate for the TEG 5000 in patients on ECMO. However, limitations in reliability were noted, and the devices may not be interchangeable when results fall outside of the reference values. © ASAIO 2021

Metrics

8 Record Views

Details

Logo image