Logo image
Effect of an UHMWPE patellar component on stress fields in the patella: A finite element analysis
Journal article   Peer reviewed

Effect of an UHMWPE patellar component on stress fields in the patella: A finite element analysis

Yeon Soo Lee, Thay Q. Lee and Joyce H. Keyak
Knee Surgery, Sports Traumatology, Arthroscopy, Vol.17(1)
2009

Abstract

Arthroplasty, Replacement, Knee Cadaver Female Finite Element Analysis Humans Knee Prosthesis Middle Aged Models, Biological Pain Patella Polyethylenes Prosthesis Design Range of Motion, Articular Stress, Mechanical polyethylene derivative ultra high molecular weight polyethylene ultra-high molecular weight polyethylene article biological model cadaver female finite element analysis human joint characteristics and functions knee arthroplasty knee prosthesis mechanical stress middle aged pain patella prosthesis
An increased stress in the patella due to the implantation of a patellar button may also be another potential source of pain in total knee arthroplasty patients. This study assessed the location inside the patella having largest stress change after implantation of an ultra high molecular polyethylene patella button. Finite elements models of the patellae before and after implantation of patellar button were created. Experimentally determined spring constants of muscles and ligaments, and patellofemoral contacting loads were applied to the models at 30°, 60°, and 90° of knee flexion. The Von Mises stress of the intact patella decreased with increased knee flexion, while that of implanted patella increased. Also, the stress range in the implanted patella was 3~9 times higher than in the intact one. The highly stressed region of the intact patella moved proximally with higher knee flexion angles, while that of the implanted model stayed near the central anterior patella. At 90° of knee flexion, the stress in the anterodistal patella increased considerably after implantation of a patella button so that the anterodistal patella may be susceptible to be painful source after the total knee replacement. © 2008 Springer-Verlag.

Metrics

8 Record Views

Details

Logo image