Late presentation of recurrent syncope after permanent pacemaker implantation due to Lead−Header malapposition
Journal article   Peer reviewed

Late presentation of recurrent syncope after permanent pacemaker implantation due to Lead−Header malapposition

Steven Mazzone, Utkarsh Kohli, Zaid A. Aziz, Andrew D. Beaser, Gaurav A. Upadhyay, Roderick Tung and Hemal M. Nayak
Indian pacing and electrophysiology journal, Vol.21(2), pp.124-127
03/01/2021
PMID: 33271274

Abstract

Lead-header malapposition Oversensing Pacemaker malfunction Recurrent syncope
Permanent pacemaker (PPM) malfunction due to electrical connection problems such as a loose set screw or lead-header malapposition is extremely rare. We present a patient with complete heart block (CHB) who had PPM malfunction and recurrent syncope, late (14 months) after initial implantation, which was caused by the ventricular lead pin disengagement from the header resulting in oversensing due to noise, pacing inhibition and recurrent syncope. PPM due to lead-header malapposition this late after device implantation has previously not been reported.