Abstract
BACKGROUNDScarce data exist on the evolution of device-related thrombus (DRT) after left atrial appendage closure (LAAC).OBJECTIVESThis study sought to assess the incidence, predictors, and clinical impact of persistent and recurrent DRT in LAAC recipients.METHODSData were obtained from an international multicenter registry including 237 patients diagnosed with DRT after LAAC. Of these, 214 patients with a subsequent imaging examination after the initial diagnosis of DRT were included. Unfavorable evolution of DRT was defined as either persisting or recurrent DRT.RESULTSDRT resolved in 153 (71.5%) cases and persisted in 61 (28.5%) cases. Larger DRT size (OR per 1-mm increase: 1.08; 95% CI: 1.02-1.15; P = 0.009) and female (OR: 2.44; 95% CI: 1.12-5.26; P = 0.02) were independently associated with persistent DRT. After DRT resolution, 82 (53.6%) of 153 patients had repeated device imaging, with 14 (17.1%) cases diagnosed with recurrent DRT. Overall, 75 (35.0%) patients had unfavorable evolution of DRT, and the sole predictor was average thrombus size at initial diagnosis (OR per 1-mm increase: 1.09; 95% CI: 1.03-1.16; P = 0.003), with an optimal cutoff size of 7 mm (OR: 2.51; 95% CI: 1.39-4.52; P = 0.002). Unfavorable evolution of DRT was associated with a higher rate of thromboembolic events compared with resolved DRT (26.7% vs 15.1%; HR: 2.13; 95% CI: 1.15-3.94; P = 0.02).CONCLUSIONSAbout one-third of DRT events had an unfavorable evolution (either persisting or recurring), with a larger initial thrombus size (particularly >7 mm) portending an increased risk. Unfavorable evolution of DRT was associated with a 2-fold higher risk of thromboembolic events compared with resolved DRT.