Abstract
The PIONEER III trial demonstrated noninferiority of 12-month target lesion failure (TLF) with the Supreme DES (Sinomed), a thin-strut cobalt-chromium, biodegradable polymer, sirolimus-eluting stent, compared with a durable polymer, everolimus-eluting (XIENCE/PROMUS) stent (DP-EES). The relative safety and effectiveness of the Supreme DES in patients with acute coronary syndromes (ACS) and those with chronic coronary syndromes (CCS) is not known.
PIONEER III was a prospective, multicenter, international, 2:1 randomized trial stratified by clinical presentation. The primary end point was TLF at 12 months (a composite of cardiac death, target vessel myocardial infarction, or ischemia-driven target lesion revascularization).
A total of 1628 patients were enrolled, including 41% of patients with ACS (unstable angina and non–ST-elevation myocardial infarction) randomized to Supreme DES (n = 441) versus DP-EES (n = 232) and 59% of patients with CCS randomized to Supreme DES (n = 645) versus DP-EES (n = 310). Patients with ACS were younger, fewer presented with less diabetes, hypertension, and previous revascularization, but more were current smokers. The primary end point of TLF (6.4% vs 4.4%; P = .1), major adverse cardiac events (8.5% vs 6.5%; P = .16), and stent thrombosis (0.4% vs 0.9%; P = .25) at 12 months were similar in the ACS and CCS groups. There was no difference in TLF at 12 months between Supreme DES and DP-EES among patients with ACS (6.6% vs 6.0%; P = .89) and those with CCS (4.5% vs 4.3%; P = .83); interaction P = .51 for TLF by clinical presentation.
Compared with the DP-EES, the Supreme DES seemed safe and effective with a similar TLF at 12 months in both patients with ACS and those with CCS.
[Display omitted]
•Supreme DES was noninferior to durable polymer EES at 12 months.•Preclinical studies show Supreme DES achieves early endothelial restoration, which may be of benefit in ACS.•Supreme DES had similar rates of TLF in patients with ACS and CCS at 12 months.•Supreme DES had similar rates of TLF to durable-polymer EES irrespective of ACS or CCS presentation.