Abstract
Background In the prehospital setting, many providers advocate for video laryngoscopy as the initial method of intubation to improve the likelihood of a successful first attempt. However, bright ambient light can worsen visualization of the video laryngoscope liquid crystal display (LCD).Case Report A patient involved in a motor vehicle accident was evaluated by an Emergency Medical Services (EMS) crew. Initial endotracheal intubation attempt using video laryngoscopy was aborted after the patient desaturated. The primary reason for the failure was poor visualization of the video laryngoscope LCD, despite attempts to block direct sunlight. Debriefing revealed that the intubating provider was wearing polarized sunglasses.Discussion Because LCDs emit polarized light, use of polarized sunglasses may cause the display to appear dark. Thus, the purpose of this Case Report is to raise awareness of a potential safety issue that is likely under-recognized by prehospital providers but can be easily avoided. SmithAJ, JackimczykK, HorwoodB, ChristensonD. © World Association for Disaster and Emergency Medicine 2019.